水中溶氧感測器(光學法) Orbisphere M1100 Oxygen Sensors
產品編號:
Sensor for the determination of dissolved oxygen (DO) in beer and de-aerated water. 分析啤酒和氣溶水中溶氧 (DO) 的感測器。
The M1100 Oxygen sensor has an unbeatable precision of 0.8ppb and a limit of detection of 0.6ppb. Such accurate measurement readings are essential to control low oxygen levels in beer. M1100 溶氧感測器具備 0.8ppb 精確度和 0.6ppb 的檢測限。如此準確的測量讀數對於控制啤酒中的低氧量極為重要。
M1100-S00 : 0 - 2000 ppb DO (dissolved oxygen)
M1100-S10 : 0 - 2000 ppb DO (dissolved oxygen)
M1100-S00H : 0 - 40 ppm DO (dissolved oxygen)
M1100-S10H : 0 - 40 ppm DO (dissolved oxygen)
The absence of membrane and electrolyte means that the sensor accuracy is unaffected by process changes or pressure shocks. Maintenance and operating costs are also reduced. The sensor has been designed to ensure mechanical robustness and resistance to CIP processes to extend operational lifetime and optimise its total cost of ownership.
The quick response time comes from a two second measurement frequency. Capable of measuring accurately at this frequency over a 12 month period.
關鍵字 : 啤酒 溶氧 分析儀, 水 含氧量 自動監測, 水 Oxygen analyzer
The M1100 Oxygen sensor has an unbeatable precision of 0.8ppb and a limit of detection of 0.6ppb. Such accurate measurement readings are essential to control low oxygen levels in beer. M1100 溶氧感測器具備 0.8ppb 精確度和 0.6ppb 的檢測限。如此準確的測量讀數對於控制啤酒中的低氧量極為重要。
M1100-S00 : 0 - 2000 ppb DO (dissolved oxygen)
M1100-S10 : 0 - 2000 ppb DO (dissolved oxygen)
M1100-S00H : 0 - 40 ppm DO (dissolved oxygen)
M1100-S10H : 0 - 40 ppm DO (dissolved oxygen)
The absence of membrane and electrolyte means that the sensor accuracy is unaffected by process changes or pressure shocks. Maintenance and operating costs are also reduced. The sensor has been designed to ensure mechanical robustness and resistance to CIP processes to extend operational lifetime and optimise its total cost of ownership.
The quick response time comes from a two second measurement frequency. Capable of measuring accurately at this frequency over a 12 month period.
關鍵字 : 啤酒 溶氧 分析儀, 水 含氧量 自動監測, 水 Oxygen analyzer
Minimal Drift and Annual Calibration
The M1100 sensor provides immediate oxygen readings with a measurement frequency of two seconds. The instrument carries consistent readings with no drift for 12 months without calibration (when utilizing standard weekly CIP processes), surpassing other optical sensors that display significant drift after only a few months in similar conditions.
Minimal Maintenance Optical Technology
The M1100 pioneered the use of luminescent technology in beverage applications, and as a result, the sensor doesn’t require the replacement of membranes or any electrolytes. Additionally, the sensor’s accuracy is una?ected by process changes or pressure shocks—further reducing maintenance. Annual maintenance is limited to just a few minutes for a zero point calibration. Chemicals are not required for this process, making the task easier and safer without reducing measurement precision.
Low Level Oxygen Measurement with Accurate ppb
The M1100-L sensor has a lower detection level of 0.6 ppb. This highly accurate instrument’s readings are essential to control low oxygen levels in beer production, to decrease unnecessary line stoppages, and to increase production uptime.
High Level Oxygen Measurement with Accurate ppm
The M1100- H sensor has a range of 0 – 40 ppm, and is ideal for use in wort applications. Even in this harsh wort environment, the instrument maintains very good accuracy and minimal drift. Only a yearly maintenance and calibration will be required in most cases.
Part Number
|
Parameter
|
Range
|
Accuracy
|
Calibration
|
---|---|---|---|---|
M1100-S00 | Oxygen | 0 - 2000 ppb DO (dissolved oxygen) | ± 0.8 ppb or 2 % whichever is greater | Single point zero calibration with standard 99.999% nitrogen (quality 50) or equivalent oxygen free gas |
M1100-S10 | Oxygen | 0 - 2000 ppb DO (dissolved oxygen) | ± 0.8 ppb or 2 % whichever is greater | Single point zero calibration |
M1100-S00H | Oxygen | 0 - 40 ppm DO (dissolved oxygen) | ± 0.02 ppm or 3 % whichever is greater | Two points at cap replacement (zero and air), one during use (air) |
M1100-S10H | Oxygen | 0 - 40 ppm DO (dissolved oxygen) | ± 0.02 ppm or 3 % whichever is greater | Two points at cap replacement (zero and air), one during use (air) |
Datasheets/Brochures | Language | Size | Date | Edition |
---|---|---|---|---|
Hach Orbisphere M1100 LDO Sensor Data Sheet LIT2635 |
English US | 246 KB | 2016-04 | Rev3 |
Soft Drink Oxygen Analyzer DOC060.53.10117 |
English US | 160 KB | 2017-01 | Sep16 |
Data Sheets | Language | Size | Date | Edition |
---|---|---|---|---|
Datasheet: Hach Orbisphere M1100 LDO Sensor DOC053.53.35143 |
English US | 1 MB | 2017-11 | Nov17 |
Orbisphere 410 Controller DOC053.53.35137 |
English US | 504 KB | 2017-12 | Nov17 |
Orbisphere 510 Multichannel Gas and Dissolved Gas Analyzer DOC053.53.35142 |
English US | 599 KB | 2017-12 | Nov17 |