水中聯氨分析儀 Polymetron 9586 sc Oxygen Scavenger Analyzer, 100 - 240 V AC
產品編號9586.99.00P2
Simple to Integrate. Simple to Operate. 整合容易, 操作簡單

Online analyzer for monitoring hydrazine and carbohydrazide oxygen scavengers. 聯氨線上分析儀

An integral part of a complete water analytics system designed for the Power industry. A broad range of product options made to work together as flexible solutions meet your unique needs. The comprehensive approach saves you time on design, installation, training, maintenance, and operation.
專為電力產業而設計的水質分析系統。 廣泛的產品選項可作為靈活的解決方案,滿足您的獨特需求。 節省設計、安裝、培訓、維護和操作的時間。

Range : 0 - 100 ppb carbohydrazide; programmable
Range : 0 - 500 ppb hydrazine; programmable

Repeatability : ± 2 % or 1 ppb whichever is greater
Response Time T90 : < 60 s

Calibration Method :
a) Zero: electrically, with hydrazine-free water or with optional zero cartridge
b) Slope: using a laboratory reference value (e.g. LCW025)

Reagent : Sample conditioning with diisopropylamine, ammonia, or diethylamine

Reference Type : Ag/AgCl electrode with electrolyte KCl 0.1 M
產品介紹
Save Time on Design
A single design source and one product platform means you spend less time searching for design files or configuring components. Create and reuse your optimal design templates.
 
Accelerate Your Installation
One source, interchangeable components, a common user interface, and one support team make installation faster and less complicated. Quickly and easily transfer user settings between oxygen scavenger analyzers.
 
Reduce Training Complexity
A single platform minimizes time required to teach and learn product operations, getting new systems in use faster.
 
Simplify Maintenance and Operation
Common menu guides reduce variability and provide step-by-step procedures for maintenance and calibration. Standard visual alerts across parameters notify operators when troubleshooting is required. The 9586 sc has a fast response time of less than 60 seconds.
Unlike traditional amperometric techniques that use two electrodes, the Hach 9586 sc oxygen scavenger analyzer uses a three-electrode design; eliminating voltage drift due to the composition of the water. Self-cleaning electrodes reduce maintenance costs and analyser downtime via PTFE beads that prevent deposits on the electrode surfaces.