電感式水中導電度感測器 3700系列 Digital Inductive Electrodeless Process Conductivity Sensor
產品編號:
Hach 3700系列感應式導電度感測器適用於嚴苛應用環境,感應式導電度之技術設計排除傳統接觸式導電度常見極化與電極coating污染的問題。
安裝時,您只需將數位感測器連結到sc系列智慧型水質數位控制主機,即可自動識別並讀取監測數據,智慧化“隨插即用”,免除複雜之接線及設定程式之步驟。
3700系列數位感應式導電度感測器量測能力: measure 200 up to 2,000,000 microSiemens/cm.
Hach’s Inductive Conductivity Sensors measure 200 up to 2,000,000 microSiemens/cm. A built-in Pt 1000 RTD compensates the measured conductivity for changes in process temperature.
The inductive sensor design eliminates polarization and electrode coating problems that commonly affect conventional contacting electrode-type conductivity sensors.
材質: Polypropylene, PFA Teflon, PEEK, PVDF
關鍵字: 水 導電度 連續監測, 水 導電度 分析儀, 水 導電度監測儀, 線上 導電度計, 水 導電度 sensor, conductivity online analyzer
安裝時,您只需將數位感測器連結到sc系列智慧型水質數位控制主機,即可自動識別並讀取監測數據,智慧化“隨插即用”,免除複雜之接線及設定程式之步驟。
3700系列數位感應式導電度感測器量測能力: measure 200 up to 2,000,000 microSiemens/cm.
Hach’s Inductive Conductivity Sensors measure 200 up to 2,000,000 microSiemens/cm. A built-in Pt 1000 RTD compensates the measured conductivity for changes in process temperature.
The inductive sensor design eliminates polarization and electrode coating problems that commonly affect conventional contacting electrode-type conductivity sensors.
材質: Polypropylene, PFA Teflon, PEEK, PVDF
關鍵字: 水 導電度 連續監測, 水 導電度 分析儀, 水 導電度監測儀, 線上 導電度計, 水 導電度 sensor, conductivity online analyzer
Hach’s Inductive Conductivity Sensors measure 200 up to 2,000,000 microSiemens/cm.
A built-in Pt 1000 RTD compensates the measured conductivity for changes in process temperature. The inductive sensor design eliminates polarization and electrode coating problems that commonly affect conventional contacting electrode-type conductivity sensors.
Wide Measuring Range
Hach’s Inductive Conductivity Sensors measure 200 up to 2,000,000 microSiemens/cm. A built-in Pt 1000 RTD compensates the measured conductivity for changes in process temperature.
Low Maintenance Design
The inductive sensor design eliminates polarization and electrode coating problems that commonly affect conventional contacting electrode-type conductivity sensors.
Versatile Mounting Styles
Sensors can be installed using a choice of four mounting styles—immersion, insertion, union, and sanitary.
Principal of Operation
Inductive conductivity sensors induce a low current in a closed loop of solution, then measure the magnitude of this current to determine the solution’s conductivity. The conductivity analyzer drives Toroid A, inducing an alternating current in the solution. This current signal flows in a closed loop through the sensor bore and surrounding solution. Toroid B senses the magnitude of the induced current which is proportional to the conductance of the solution. The analyzer processes this signal and displays the corresponding reading.
Withstands Harsh Environments
The inductive sensor is available in sanitary (CIP) flange style and convertible styles in PFA®, polypropylene, PEEK®, and PVDF material. Select sensors can withstand high pressures and temperatures.