數位電感式水中導電度感測器 Digital Convertible Mounting Style Inductive, Electrodeless Conductivity Sensor, PFA Body Material
產品編號D3728E2T
PFA, Convertible Mounting Style, 6m (20 ft) Analog Cable, Electrodeless Conductivity Sensor, includes 6120800 Digital Gateway for Connecting to sc Controller and a 6122400 1m (3 ft) Digital Extension Cable

Range 量測範圍 : 200 µS/cm - 2000 mS/cm
Accuracy 準確度 : ± 0.01 % of reading, all ranges
Flow 流速 : Rate 3 m (10 ft.) per second, maximum
Operating Temperature Range 操作溫度 : -10 - 200 °C
Sensor Type 感測器類別 : Digital 數位

Applications 應用 :
Membrane Protection
Nutrient Monitoring
Secondary Treatment
Ultrapure Water

關鍵字 : 水中 導電度 線上分析儀, 水中 導電度監測儀, 連續監測 水中導電度, 線上 水中 導電度計, online conductivity analyzer,
產品介紹
Wide Measuring Range  量測範圍寬廣
Hach’s Inductive Conductivity Sensors measure 200 up to 2,000,000 microSiemens/cm. A built-in Pt 1000 RTD compensates the measured conductivity for changes in process temperature.
 
Low Maintenance Design  電感式感測器設計 減低日常維護工作量
The inductive sensor design eliminates polarization and electrode coating problems that commonly affect conventional contacting electrode-type conductivity sensors.
 
Versatile Mounting Styles  靈活安裝方式
Sensors can be installed using a choice of four mounting styles—immersion, insertion, union, and sanitary.
 
Principal of Operation  操作原理
Inductive conductivity sensors induce a low current in a closed loop of solution, then measure the magnitude of this current to determine the solution’s conductivity. The conductivity analyzer drives Toroid A, inducing an alternating current in the solution. This current signal flows in a closed loop through the sensor bore and surrounding solution. Toroid B senses the magnitude of the induced current which is proportional to the conductance of the solution. The analyzer processes this signal and displays the corresponding reading.
 
Withstands Harsh Environments 耐受嚴苛環境
The inductive sensor is available in sanitary (CIP) flange style and convertible styles in PFA®, polypropylene, PEEK®, and PVDF material. Select sensors can withstand high pressures and temperatures.