水中導電度線上感測器(鍋爐水/冷凝水) Stainless Steel Digital Contacting Conductivity Sensor for High Conductivity, k=5.0, Boiler/Condensate
產品編號:3444D8A
Ultimate accuracy from ultra-pure to high conductivity applications. 滿足從超純水到高導電度應用的準確度。
Nominal Cell Constant (K) = 5.0, 3/4 inch male NPT threaded 316 stainless steel body, 6m (20 ft) Analog Cable, Titanium Electrode
Range 量測範圍 : 0 - 10000 µS/cm
Repeatability 重複性 : ±0.5 % of reading
Accuracy 準確度 : ±2 % of reading above 200 µS/cm
Cell Constant k : 5
Response Time T90 : 30 s of step change
Sensitivity : ±0.5 % of reading
Sensor Type : Analog
Temperature Measurement : -20 - 200 °C
Temperature Sensor 溫度感測器 : PT1000 RTD
Installation style 安裝方式 : Boiler 鍋爐水 / Condensate 冷凝水
*使用Hach SC系列 digital controllers 數位控制主機進行監控
關鍵字:純水 導電度 線上監測儀, 純水 導電度 線上分析儀, 冷凝水 導電度 線上監測儀, online conductivity analyzer,
Nominal Cell Constant (K) = 5.0, 3/4 inch male NPT threaded 316 stainless steel body, 6m (20 ft) Analog Cable, Titanium Electrode
Range 量測範圍 : 0 - 10000 µS/cm
Repeatability 重複性 : ±0.5 % of reading
Accuracy 準確度 : ±2 % of reading above 200 µS/cm
Cell Constant k : 5
Response Time T90 : 30 s of step change
Sensitivity : ±0.5 % of reading
Sensor Type : Analog
Temperature Measurement : -20 - 200 °C
Temperature Sensor 溫度感測器 : PT1000 RTD
Installation style 安裝方式 : Boiler 鍋爐水 / Condensate 冷凝水
*使用Hach SC系列 digital controllers 數位控制主機進行監控
關鍵字:純水 導電度 線上監測儀, 純水 導電度 線上分析儀, 冷凝水 導電度 線上監測儀, online conductivity analyzer,
High Performance Design 高效能設計
These sensors are manufactured to exacting tolerances using high quality, rugged materials for demanding applications including ultra-pure water, clean-in-place (CIP), and boiler/condensate monitoring. Each sensor is tested to determine its unique, absolute four-digit cell constant. Also, each sensor has a PT1000 RTD temperature element built into its tip for exceptionally fast response to changes in temperature with ±0.1 ºC accuracy.
These sensors are manufactured to exacting tolerances using high quality, rugged materials for demanding applications including ultra-pure water, clean-in-place (CIP), and boiler/condensate monitoring. Each sensor is tested to determine its unique, absolute four-digit cell constant. Also, each sensor has a PT1000 RTD temperature element built into its tip for exceptionally fast response to changes in temperature with ±0.1 ºC accuracy.
Resistivity and Conductivity Measurement Capability 電阻和導電度測量準確度
These enhanced performance sensors measure from theoretically pure water (0.057 µS/cm or 18.2 MΩ) up to 200,000 µS/cm. Hach’s sc Digital Controllers accept multiple digital sensor inputs, and can be user-set to measure conductivity, resistivity,TDS, salinity, or one of six calculated measurements.
These enhanced performance sensors measure from theoretically pure water (0.057 µS/cm or 18.2 MΩ) up to 200,000 µS/cm. Hach’s sc Digital Controllers accept multiple digital sensor inputs, and can be user-set to measure conductivity, resistivity,TDS, salinity, or one of six calculated measurements.
Versatile Mounting Styles 靈活安裝模式
High pressure and high temperature sensors—Are designed for monitoring boiler water and condensate in return lines. They have 316 stainless steel electrodes and threaded bodies (3/4-inch male NPT). They can be fastened into a boiler wall using a 3/4-inch weldolet or mounted into a process line using a standard 3/4-inch stainless pipe tee.
High pressure and high temperature sensors—Are designed for monitoring boiler water and condensate in return lines. They have 316 stainless steel electrodes and threaded bodies (3/4-inch male NPT). They can be fastened into a boiler wall using a 3/4-inch weldolet or mounted into a process line using a standard 3/4-inch stainless pipe tee.
Full-Featured “Plug and Play” Hach sc Digital Controllers 與Hach 所有SC系列數位控制主機相容, 即插即用
There are no complicated wiring or set up procedures with any Hach sc controller. Just plug in any combination of Hach digital sensors and it's ready to use—it’s “plug and play.”
There are no complicated wiring or set up procedures with any Hach sc controller. Just plug in any combination of Hach digital sensors and it's ready to use—it’s “plug and play.”