電感式水中導電度感測器 Digital Convertible Mounting Style Inductive (Electrodeless) Conductivity Sensor with Peek Body Material
產品編號D3727E2T
PEEK, Convertible Mounting Style, 6m (20ft) Analog Cable, Electrodeless Conductivity Sensor, includes 6120800 Digital Gateway for Connecting to sc Controller and a 6122400 1m (3ft) Digital Extension Cable

Range 量測範圍 : 200 µS/cm - 2000 mS/cm
Accuracy 準確度 : ± 0.01 % of reading, all ranges
Flow 流速 : Rate 3 m (10 ft.) per second, maximum
Body Material 材質 : PEEK
Cable Length : 22 ft
Operating Temperature Range 操作溫度範圍 : -10 - 200 °C
Temperature Sensor 溫度感測器 : Temperature Compensator Pt 1000 RTD

Sensor Cable 感測器電纜 : Polypropylene and PVDF Sensors:
5 conductor (plus two isolated shields) cable with XLPE (cross-linked polyethylene) jacket; rated to 150 °C (302ºF); 6 m (20 ft.) long

Sensor Type 感測器類別 : Digital 數位

*使用Hach sc 系列 controllers 水質控制主機進行監控

Applications 應用 :
Membrane Protection
Nutrient Monitoring
Secondary Treatment
Ultrapure Water

關鍵字 : 水中 導電度計, 水中 導電度 線上監測儀, 水中 導電度 自動分析儀, conductivity analyzer, 連續 監測 水中 導電度
產品介紹
Wide Measuring Range 量測範圍寬廣
Hach’s Inductive Conductivity Sensors measure 200 up to 2,000,000 microSiemens/cm. A built-in Pt 1000 RTD compensates the measured conductivity for changes in process temperature.
 
Low Maintenance Design  電感式導電度感測設計 相較於傳統導電度計  大幅減低日常維護工作
The inductive sensor design eliminates polarization and electrode coating problems that commonly affect conventional contacting electrode-type conductivity sensors.  電感式感測器設計免除傳統接觸式電極型導電度感測器的極化和電極塗層問題
 
Versatile Mounting Styles 靈活安裝方式
Sensors can be installed using a choice of four mounting styles—immersion, insertion, union, and sanitary.
 
Principal of Operation
Inductive conductivity sensors induce a low current in a closed loop of solution, then measure the magnitude of this current to determine the solution’s conductivity. The conductivity analyzer drives Toroid A, inducing an alternating current in the solution. This current signal flows
in a closed loop through the sensor bore and surrounding solution. Toroid B senses the magnitude of the induced current which is proportional to the conductance of the solution. The analyzer processes this signal and displays the corresponding reading.
 
Withstands Harsh Environments  耐受嚴苛環境
The inductive sensor is available in sanitary (CIP) flange style and convertible styles in PFA®, polypropylene, PEEK®, and PVDF material. Select sensors can withstand high pressures and temperatures.