水中pH值感測器 Hach pHD Online Process pH Sensor - pH Sensor for Clean Water
產品編號:DPD1R1
Clean Water Online Process pH Sensor with Integrated Digital Electronics for "Plug and Play" with Hach Digital SC Controllers - pHD Technology, Glass pH Electrode, Ryton Housing, Convertible Mount, 10 m Cable. 整合數位電子元件的淨水 pH 感測器,用於“即插即用”的 Hach sc系列數位控制器 - 具有 pHD 技術、玻璃 pH 電極、Ryton 外殼、可轉換的安裝座、10公尺電纜
This instrument connects to Claros, Hach’s innovative Water Intelligence System, enabling you to seamlessly connect and manage instruments, data, and process – anywhere, anytime. The result is greater confidence in your data and improved efficiency in your operations.
Exceptional Process pH Sensor Performance with the Differential Electrode pHD Measurement Technique 差分電極 pHD 測量技術
Lower Maintenance Needs with the Double Junction Salt Bridge
Extended Working Life with the Replaceable Salt Bridge/Protector
Reliability with Built-in Encapsulated Preamp
Electrode Type 探頭種類 : General Purpose
Material Ground Electrode : Titanium
Range 量測範圍 : -2.0 - 14.0 pH
Repeatability 重複性 : ± 0.05 pH
Calibration Method 校正方法 : Two point automatic, one point automatic, two point manual, one point manual.
Operating Temperature Range : -5 - 95 °C
Temperature Compensation 溫度補償 : automatic with NTC 300 Ω thermistor, or manually fixed at a user-entered temperature,
additional selectable temperature correction factors (ammonia, morpholine, or user-defined pH/°C linear slope) available for pure water automatic compensation 0.0 - 50 °
Communication 通訊 : MODBUS
Applications 應用 : Membrane Protection, Nutrient Monitoring, Secondary Treatment, Ultrapure Water
*使用Hach sc系列 Digital Controllers 數位水質控制主機進行控制
關鍵字: 水 線上 pH 線上分析儀, 水 pH 濃度 監測, 線上 pH 計, pH sensor, water pH analyzer, 水中 酸鹼值分析儀
This instrument connects to Claros, Hach’s innovative Water Intelligence System, enabling you to seamlessly connect and manage instruments, data, and process – anywhere, anytime. The result is greater confidence in your data and improved efficiency in your operations.
Exceptional Process pH Sensor Performance with the Differential Electrode pHD Measurement Technique 差分電極 pHD 測量技術
Lower Maintenance Needs with the Double Junction Salt Bridge
Extended Working Life with the Replaceable Salt Bridge/Protector
Reliability with Built-in Encapsulated Preamp
Electrode Type 探頭種類 : General Purpose
Material Ground Electrode : Titanium
Range 量測範圍 : -2.0 - 14.0 pH
Repeatability 重複性 : ± 0.05 pH
Calibration Method 校正方法 : Two point automatic, one point automatic, two point manual, one point manual.
Operating Temperature Range : -5 - 95 °C
Temperature Compensation 溫度補償 : automatic with NTC 300 Ω thermistor, or manually fixed at a user-entered temperature,
additional selectable temperature correction factors (ammonia, morpholine, or user-defined pH/°C linear slope) available for pure water automatic compensation 0.0 - 50 °
Communication 通訊 : MODBUS
Applications 應用 : Membrane Protection, Nutrient Monitoring, Secondary Treatment, Ultrapure Water
*使用Hach sc系列 Digital Controllers 數位水質控制主機進行控制
關鍵字: 水 線上 pH 線上分析儀, 水 pH 濃度 監測, 線上 pH 計, pH sensor, water pH analyzer, 水中 酸鹼值分析儀
The smart choice for accurate and reliable online process pH measurement
Exceptional Process pH Sensor Performance with the Differential Electrode pHD Measurement Technique
This field-proven technique uses three electrodes instead of the two normally used in conventional pH sensors. Process and reference electrodes measure the pH differentially with respect to a third ground electrode. The end result is unsurpassed measurement accuracy, reduced reference junction potential, and elimination of sensor ground loops. These process pH sensors provide greater reliability, resulting in less downtime and maintenance.
Lower Maintenance Needs with the Double Junction Salt Bridge
The double junction salt bridge creates a barrier to contamination which minimizes the dilution of the internal standard cell solution. The result is lower maintenance needs and a longer time period between calibrations.
Extended Working Life with the Replaceable Salt Bridge/Protector
The unique, replaceable salt bridge holds an extraordinary volume of buffer to extend the working life of the sensor by protecting the reference electrode from harsh process conditions. The salt bridge simply threads onto the end of the sensor if replacement is needed.
Reliability with Built-in Encapsulated Preamp
Encapsulated construction protects the sensor’s built-in preamp from moisture and humidity, ensuring reliable sensor operation. The preamp in the pHD analog sensor produces a strong signal, enabling the sensor to be located up to 1000 m (3280 ft.) from the analyzer.
Innovative Technology
The former GLI, now a Hach Company brand, invented the Differential Electrode Technique for pH measurement in 1970. The pHD™ sensor series (U.S. Patent Number 6395158B1, dated May 28, 2002) takes this field-proven technology to a new level.