水中生化需氧量BOD分析電極 Intellical™ LBOD101 Luminescent/Optical Dissolved Oxygen Sensor for BOD Measurements
產品編號LBOD10101
A Breakthrough in BOD Measurement!
Hach LBOD Intellical™ probe uses patented LDO® technology to deliver superior confidence in results.
LBOD101 探頭測量 BOD5 測試應用中 BOD 瓶中生化需氧量的溶氧。 配備自動壓力感測模組。 iButton功能追蹤使用天數並提醒感測帽元件的剩餘壽命。
Intellical™ LBOD101 is the latest embodiment of Hach’s patented Luminescent Dissolved Oxygen (LDO) technology. The LBOD101 is a digital, luminescent dissolved oxygen (LDO) sensor with a replaceable, integrated stirring system. The LBOD101 measures the dissolved oxygen for BOD (Biochemical Oxygen Demand) determination in BOD bottles in BOD5 testing applications. The LBOD101 probe is used for initial and final dissolved oxygen measurements. The LBOD101 is fitted with an automatic pressure sensor module. The DO sensing cap comes with an iButton to track days in use and remind remaining life of the sensing cap element.

Range 量測範圍 : 0.05 - 20.0 mg/L
1 - 200% saturation
Resolution : 0.01/0.1

*使用 Hach HQ系列主機進行分析

關鍵字 : 水中 BOD 測試 探頭, 水中 生化需氧量 檢測
產品介紹
BOD (Biochemical Oxygen Demand) Method Range Reagent Set(s) USEPA Compliant?
  10360 0.05 - 20.00 mg/L BOD View  
 
The Hach Intellical™ LBOD101 optical dissolved oxygen probe improves confidence in the BOD5 testing procedure by reducing variance and maintenance while increasing stability, accuracy, precision, and efficiency (tests per unit time).
 
The Intellical™ LBOD101 probe has no membranes to foul or replace, no electrolyte solution to replenish, and fast, reliable 2-point calibration that stabilizes quickly. Additionally, there is no 30-minute polarization and warm-up time with the LBOD101 probe.
 
Intellical™ LBOD101 low maintenance requirements keeps lab down-time to a minimum
The Intellical™ LBOD101 has no membranes to replace (lower cost of ownership), features a replaceable stirrer assembly and a long-life patented magnetic-based motor that is also replaceable.
 
Intellical™ digital probes provide ultimate traceability in measurement history
Stored time and date stamp for each measurement, operator and sample ID, calibration history, parameter, and probe serial number.
 
Intellical™ digital probes can be moved between meters without the need to re-calibrate or re-enter measurement settings
Ideal for a multi-user environment with multiple HQD series Laboratory and Portable meters.
 
Guides Language Size Date Edition
HQD Digital Meters and Intellical™ Probes Data Sheet
LIT2599
English US 809 KB 2017-11 Rev9
Measuring Chloride in Power Applications
DOC040.53.10132
English US 707 KB 2018-02 Jan18
Manuals Language Size Date Edition
LBOD101 Probe Basic User Manual-Multiple Languages
DOC022.97.80025
English US, French, Spanish, Portuguese, Chinese, Japanese, Korean, Thai 857 KB 2014-09 Ed 3
User Manual for BOD LDO® Probe, Model LBOD10101
DOC022.53.80025
English US 462 KB 2014-09 Ed 3
Methods/Procedures Language Size Date Edition
Oxygen Demand, Biochemical, Dilution Method 10360, LBOD Measurement
DOC316.53.01242