水中硝酸鹽離子選擇分析電極 Intellical™ ISENO318101 Nitrate (NO3-) Ion Selective Electrode (ISE) with Calibration Reagents Pack
產品編號: ISENO3181AP
Hach多參數數位水質分析儀HQd 系列目前可檢測11 種參數,具有標準型、堅固型以及Ultra 等多種電極選擇。極大測量靈活性,更換電極非常簡單,且可自動識別電極。各種IntelliCAL智慧型電極,可測量溫度、pH值、導電度、溶氧(LDO),LBOD,ORP、鈉、銨、氨、氟、硝酸鹽、氯等。
The Intellical™ Nitrate ISE Analysis Package contains all you need to start measuring nitrate (NO3-) with your HQD meter instrument: The Intellical™ ISENO3181 Nitrate ISE with 1 m cable, 500 mL bottles of 1 mg/L, 10 mg/L and 100 mg/L of Hach nitrogen-nitrate standard solutions, pack of 100 Hach nitrate ionic strength adjustor (ISA) Powder Pillows.
Parameter : Nitrate
Method Type : Laboratory: Non-Refillable Dritek Gel Reference Element
Measuring range : 0.1 mg/L (7 x 10-6 M) - 14 g/L (1 M) NO3-N
*Utilizes a unique solid-state sensor technology that eliminates costly membrane replacements.
*Requires virtually no maintenance
*Fast, stable, and accurate results
*Ultimate traceability in measurement history
*Can be moved between meters without the need to re-calibrate or re-enter measurement settings
*使用 HQd系列水質分析主機(Meters)進行分析
關鍵字 : 實驗室 水中 硝酸鹽 濃度 分析儀, 水中 硝酸鹽 分析探頭
The Intellical™ Nitrate ISE Analysis Package contains all you need to start measuring nitrate (NO3-) with your HQD meter instrument: The Intellical™ ISENO3181 Nitrate ISE with 1 m cable, 500 mL bottles of 1 mg/L, 10 mg/L and 100 mg/L of Hach nitrogen-nitrate standard solutions, pack of 100 Hach nitrate ionic strength adjustor (ISA) Powder Pillows.
Parameter : Nitrate
Method Type : Laboratory: Non-Refillable Dritek Gel Reference Element
Measuring range : 0.1 mg/L (7 x 10-6 M) - 14 g/L (1 M) NO3-N
*Utilizes a unique solid-state sensor technology that eliminates costly membrane replacements.
*Requires virtually no maintenance
*Fast, stable, and accurate results
*Ultimate traceability in measurement history
*Can be moved between meters without the need to re-calibrate or re-enter measurement settings
*使用 HQd系列水質分析主機(Meters)進行分析
關鍵字 : 實驗室 水中 硝酸鹽 濃度 分析儀, 水中 硝酸鹽 分析探頭
Utilizes a unique solid-state sensor technology that eliminates costly membrane replacements.
Traditional PVC-membrane ISEs require membrane replacement every 2-3 months. The Intellical ISENO3181 Nitrate ISE’s unique solid-state sensor technology eliminates the need for frequent membrane replacement by using a solid gel ion exchange behind the ion-selective membrane versus the liquid ion exchange used in most ISEs. This design allows users to get up to 2 years of life from their ISE without membrane replacement – saving the time and costly expense of frequently exchanging membranes.
Traditional PVC-membrane ISEs require membrane replacement every 2-3 months. The Intellical ISENO3181 Nitrate ISE’s unique solid-state sensor technology eliminates the need for frequent membrane replacement by using a solid gel ion exchange behind the ion-selective membrane versus the liquid ion exchange used in most ISEs. This design allows users to get up to 2 years of life from their ISE without membrane replacement – saving the time and costly expense of frequently exchanging membranes.
Requires virtually no maintenance
The solid-state sensor design allows for DRY storage of the ISE without a shelf life. Also, the probe does not require the electrolyte to be refilled or membrane replacement.
The solid-state sensor design allows for DRY storage of the ISE without a shelf life. Also, the probe does not require the electrolyte to be refilled or membrane replacement.
Fast, stable, and accurate results
Intellical digital probes lock on the result when the measurement is stable, removing the guesswork in having a moving number. Additionally, the probe’s large annular reference junction provides maximum surface area between sample and reference electrolyte for fast measurement stability. And the solid-state sensor design doesn’t allow the membrane to flex or undergo surface area changes, leading to more stability in results over a variety of samples.
Intellical digital probes lock on the result when the measurement is stable, removing the guesswork in having a moving number. Additionally, the probe’s large annular reference junction provides maximum surface area between sample and reference electrolyte for fast measurement stability. And the solid-state sensor design doesn’t allow the membrane to flex or undergo surface area changes, leading to more stability in results over a variety of samples.
Ultimate traceability in measurement history
Stored time and date stamp for each measurement, operator and sample ID, calibration history, parameter, and probe serial number
Stored time and date stamp for each measurement, operator and sample ID, calibration history, parameter, and probe serial number
Can be moved between meters without the need to re-calibrate or re-enter measurement settings
Ideal for a multi-user environment with multiple HQD series Laboratory and Portable meters
Ideal for a multi-user environment with multiple HQD series Laboratory and Portable meters
Nitrate | Method | Range | Reagent Set(s) | USEPA Compliant? |
---|---|---|---|---|
8358 Direct Measurement ISE | 0.1 - 100.0 mg/L NO3-N | View | ||
8358 Direct Measurement ISE | 0.1 - 14,000 mg/L NO3-N | View | ||
8359 Direct Measurement ISE | 0.04 - 4.00 mg/L NO3 -N |
View |